Do not connect a real analyzer until

  • The vendor interface manual and sample messages have been reviewed.
  • Analyzer test codes, units, flags, barcode field, and transport settings are mapped.
  • The analyzer’s network access is limited to the laboratory VLAN.
  • OIE administration is HTTPS/VPN restricted.
  • OIE channel exports and database backups are in place.
  • Lab staff have completed reconciliation and downtime procedures.
  • Results are verified in a monitored pilot before routine clinical use.
Bidirectional orders/worklists should be enabled only after the analyzer confirms its supported order interface.